Startup and Operation

Model 2240

Teledyne Analytical Instruments

6

2.2 Setup

Insert the probe cable into the probe and the other end into the

Controller. The cable has distinct connections that will only mate in the
proper manner.

2.3 Startup

Power-up the Model 2240 by pressing and holding the ON/OFF

button until the Controller LCD display indicates an operational
message.

Note: Only power-up the instrument in a hydrogen-free

environment.

After power is on the instrument automatically enters an

Initialization mode that lasts approximately ten minutes. During this
time the LCD displays a countdown to completion and the lighted
indicator closest to the Probe Tip is orange (See Figure 2-2) until the
following operations complete:

Heat the Wide Range Sensor® to operating temperature.

Perform system self-test.

Zero the sensor reading.

Figure 2-2: Startup LED

Upon successful completion the instrument automatically switches

to Normal operation and the Probe tip LED turns green. If an error is
detected the instrument will display an Error Status (see Section 2.8).