Handheld Hydrogen Leak Detector Model 2240

Teledyne Analytical Instruments

Page 29 of 30

11

Apply
0.100%H2
Settle

Checking sensor
temperature.

12

Apply
0.100%H2
Wait xxxx

Wait for sensor reading to
stabilize until xxxx = 0.

13

Apply
0.100%H2
Finding
Average

Measuring sensor response
to test gas.
Visually verify that 1-2 yellow
LED’s are lit up in the probe
tip. If not the unit needs
factory calibration

14

Apply
2.000%H2
Continue?

With the Calibration Cup,
apply 2.0% hydrogen to the
Probe sensor. The Probe
Tip LED will remain Red.
Three yellow LEDs in the
LED Bar Graph Array will
turn on. Press ENTER.

15

Apply
2.000%H2
In Progress

2.0% Verify Test starts.

16

Apply
2.000%H2
Settle

Checking sensor
temperature.

17

Apply
2.000%H2
Wait xxxx

Wait for sensor reading to
stabilize until xxxx = 0.

18

Apply
2.000%H2
Finding
Average

Measuring sensor response
to test gas.

19

Enter Date:
1.0000 M

Enter the current month (1-
12) using the

▲ (up arrow)

and

▼ (down arrow) keys.

20

Enter Date:
1.0000 D

Enter the current day (1-31)
using the

▲ (up arrow) and

▼ (down arrow) keys.