Handheld Hydrogen Leak Detector Model 2240

Teledyne Analytical Instruments

Page 28 of 30

Table 1 - Verify Procedure

Step Display

User

response

1

Verify Sensor

Press ENTER

2

Verify Sensor
Continue?

Press ENTER to Verify
sensor, L to exit.

3

Verify Sensor
In Progress

Verify Test begins.

4

Apply
0.000%H2
Continue?

With the Calibration Cup that
accompanies the MODEL
2240, apply hydrogen-free,
ultra-zero air to the Probe
sensor. The Probe Tip LED
will remain Green. Press
ENTER.

5

Apply
0.000%H2
In Progress

0% Verify Test starts.

6

Apply
0.000%H2
Settle

Checking sensor
temperature.

7

Apply
0.000%H2
Wait xxxx

Wait for sensor reading to
stabilize until xxxx = 0.

8

Apply
0.000%H2
Finding
Average

Measuring sensor response
to test gas.

9

Apply
0.100%H2
Continue?

With the Calibration Cup,
apply 0.1% hydrogen to the
Probe sensor. The Probe
Tip LED will change from
Green to Red. One (or two)
yellow LEDs in the LED Bar
Graph Array will turn on.
Press ENTER.

10

Apply
0.100%H2
In Progress

0.1% Verify Test starts.