Handheld Hydrogen Leak Detector Model 2240

Teledyne Analytical Instruments

Page 25 of 30

10. Troubleshooting

Symptom

Possible Cause

Action

Sensor Error

The probe is
disconnected from
the controller.

Turn off the instrument
and verify that the probe
is properly connected to
the controller.

Error 88

Faulty capacitor

Turn off the instrument.

Error 40

The PCB
temperature is too
high.

Turn off the instrument.
Let it cool.

Error 20

The sensor
temperature is
incorrect.

Turn off the instrument.

Battery LED is
red

The battery is very
low.

Charge the battery
completely. See
Section 8.

Unit won’t turn
on

The battery is
dead.

Charge the battery
completely. See
Section 8.

If the recommended action does not solve the problem, the
MODEL 2240 should be returned to the factory for repair.