Trace Nitrogen in Argon Analyzer

Maintenance

To replace the PMT:

1. Remove the AC power cord from the unit.

2. When removing the PMT cover, the optical filter dislodges from

its position. Remove the PMT cover, carefully catching the
optical filter with gloved hands to avoid getting any fingerprints
on the filter.

3. Remove the PMT.

4. Install the new PMT.

5. Clean the glass surface of the PMT, removing any fingerprints

or dust.

6. Replace the optical filter into the window opening with the

mirror-side out. Hold it in place.

7. Replace the cover on the PMT. Take care to keep the filter and

window lined up with the vertical grid visible within the tube.

5.4.7 High Voltage Plate Adjustment

If the plate voltage to the PMT is too high, a non-linearity problem

will arise where gases in the midrange of the analyzer read higher
than normal. If the plate voltage to the PMT is too low, the analyzer
will run out of span adjustment on the front panel control. For this
reason, the plate voltage to the PMT is critical.

Each new analyzer is tested with a range of gases to determine the

optimum high plate voltage for the PMT. This voltage is adjustable
through the potentiometer, labeled 1 P28 Plate Adjust on the rear panel
of the analyzer. See Figure 1-2.

Adjust this voltage very carefully to eliminate non-linearity

problems. It is normally only adjusted if you run out of span or zero
adjustment on the front panel controls.

5.4.8 High Voltage Transformer

One of the most common failures of the analyzer is an open

secondary winding in the high-voltage transformer. This transformer,
powered by the 255 Hz power supply, supplies ionization voltage to the
analytical cell. The value of high voltage coming from the transformer is
difficult to measure, but an accurate reading is not necessary. To check
the transformer: