5 gas connections, 5 gas connections 22, Figure 3-2: electrical connections – Teledyne 2120 - Trace Nitrogen in Argon Analyzer User Manual

Page 32

background image

Installation Model

2120

4. Connect the 4-20 mA analog recorder output terminals to a

recorder, if desired. The recorder output terminals are located
on the terminal strip on the back of the analyzer as shown in
Figure 3-2.

Figure 3-2: Electrical Connections

3.5 Gas Connections

CAUTION:

THIS INSTRUMENT IS NOT DESIGNED TO HANDLE
HAZARDOUS GASES.

OZONE EMISSIONS CAN OCCUR IF AIR OR
MOISTURE ARE IONIZED IN THE SAMPLING
SYSTEM. THE SAMPLE PATH MUST BE
RELATIVELY FREE OF AIR AND/OR MOISTURE
BEFORE APPLYING THE IONIZING VOLTAGE.

KEEP THE SERIES 1200 POWER OFF UNTIL THE
ANALYZER IS FULLY PURGED.

Note: Air leaking into the sampling system will cause erratic or

unsatisfactory analyzer operation. Even if air is admitted
into the system for only a few minutes, you must purge the
regulator and the system for at least 1 hour before the
readings stabilize. See Section 3.6.

22

Teledyne Analytical Instruments