Maintenance Model

2120

Table 5-1 summarizes the recommended routine maintenance

required for the Model 2120.

Table 5-1: Routine Maintenance Schedule

Maintenance Task

Time Schedule

Calibrate Every

1-2

weeks

Check sample and reference gas flow rates

Every 1-2 weeks

Check for excessive accumulation of dirt

Weekly

Check gas lines for leaks

Yearly

5.3 Troubleshooting

WARNING: HIGH VOLTAGE. ELECTROCUTION HAZARD. NEVER

OPERATE THE ANALYZER WITHOUT THE COVER
SECURED IN PLACE. DO NOT TAMPER WITH OR
TRY TO ADJUST THE ANALYZER UNLESS YOU ARE
THOROUGHLY KNOWLEDGEABLE OF THE
INSTRUMENT AND ITS OPERATION. ALWAYS TURN
OFF POWER AND DISCONNECT THE AC POWER
CORD BEFORE SERVICING.

Intervention should remain limited to the replacement of defective

parts.

Before consulting the troubleshooting guide in Table 5-2, ensure

that the sample flow is constant and vent pressure is at atmospheric
pressure.

Table 5-2 provides you with information on problems that may

occur with the Model 2120 Analyzer. Also included are their
probable causes and suggested remedies to the problem. Standard
hand tools and a digital multimeter are the only equipment required for
most procedures.

Note: If you cannot fix the problem using these guidelines,

contact Teledyne Customer Service (626) 961-9221 and be
ready to give details regarding your installation, including
the analyzer's serial number found on the identification
plate located on the back panel.