3.8 Installation Checklist
After installing the system, and prior to powering up the system,
verify the following items:
1. Each gas line has been purged for at least 1 hour prior to
connection to the analyzer inlets.
2. There are no leaks at the sample inlet gas connection to the
instrument.
3. There are no leaks between the span gas cylinder and the
analyzer span gas inlet port when span gas is flowing to the
analyzer.
4. There are no leaks between the zero gas source and the
analyzer zero gas inlet port when zero gas is flowing to the
analyzer.
5. The vent tube connection has no leaks, has an unobstructed
path to atmosphere, and is at least 1/4-in (0.64-cm) OD.
6. All electrical connections have been made properly and the
Power Entry module is set for the appropriate power from your
facility.