Trace Nitrogen in Argon Analyzer

Introduction

Figure 1-2: Model 2120 Rear Panel

Recorder Contacts — A standard barrier terminal connection is

installed for connecting a recorder or other external meter to the
analyzer.

Test Point Contacts —Provides an mVDC output signal

connector for a DC multimeter and is used during calibration.

Overrange Alarm Contacts* — Normally open and normally

closed dry contacts for indication of impurities at levels too
high for analyzer's range of detection. Rated for 0.5 A at 30
VAC.

Range Identification Contacts* — These contacts are closed

when the lower range is selected on the front panel of the
analyzer and open for the high range. Rated for 2 A at 30
VAC.

Sample Vent Port — 1/4-in (0.64 cm) Swagelok compression

fitting marked VENT. Used to connect the sample outlet line to
the vent system.

1 P28 Plate Adjustment — A 10-turn, 10 K variable plate

voltage adjustment for the high voltage applied to the plates of
the PMT by the high-voltage DC power supply.

Identification Plate — This plate identifies the model #, serial

#, range, etc.

Test Points — Test jacks are installed to test the voltage outputs

of the amplifier board, the low-voltage power supply, and the
255 Hz power supply.

*Note: A fuse must be used for any circuit attached to the

overrange alarm or range indication contacts

.