2

012-10115

Test Method Description

The dissolved CO

2

ezSample

test method employs the

a caustic titrant with pH indicator method.

1,2

Results

are expressed in ppm (mg/L) carbon dioxide as CO

2

.

Procedure in the event of sulfide concentrations > 0.4 ppm:

Sulfide will not interfere up to 0.4 ppm. However if the sulfide concentration is > 0.4 ppm, the
following formula is used to calculate the volume of A-1905 Neutralizer Solution that should be added
to 20 mL of the sample prior to performing the Test Procedure:

mL of A-1905 Solution = ppm sulfide ÷ 10

.

NOTE: Because the ampoules have nonlinear scales, the accuracy of the ezSample field titrator kit varies with the analyte
concentration. At the low end of the test range, the accuracy is

± 5%. At the high end of the range, the accuracy falls to ± 20%.

References

1. Method

4500—CO

2

C. APHA Standard Methods, 20th ed., p. 4-31, (1998).

2

Total and Dissolved Carbon Dioxide In Water, Test Method E. ASTM D. 513-82.