beautypg.com

Mounting the sensor, Sensor specifications, Technical support – PASCO CI-6520A Magnetic Field Sensor User Manual

Page 3

background image

®

C I - 6 5 2 0 A

0 1 2 - 0 6 9 2 8 C

M o u n t i n g t h e S e n s o r

3

The RANGE SELECT slide switch on the top of the sensor selects the magnetic field range:

100X, range ±10 gauss; resolution, 50 milligauss (0.050 G). Variations in the earth’s magnetic field and small fields around
power cords can be detected. (NOTE: Using the EM-8652 Zero Gauss Chamber when zeroing the sensor on this setting is rec-
ommended.)

10X, range ±100 gauss; resolution, 50 milligauss (0.05 G). Use this range with current-carrying solenoids and Helmholtz coils.

1X, range ±1000 gauss, resolution, 500 milligauss (0.5 G). Use this range for fields typically found around strong permanent
magnets.

Mounting the Sensor

The included Sensor Mounting Stud can be screwed
into the 1/4-20 threaded hole located on the bottom
of the sensor. Use the Sensor Mounting Stud to
mount the sensor to a support rod, for example.

Sensor Specifications

(*1 gauss = 0.0001 tesla or 10

-4

T)

1

For example, if the value reading on the ±100 gauss range is 60 gauss, the accuracy is ±6 gauss (10% of the reading).

Technical Support

For assistance with any PASCO product, contact PASCO at:

For the latest information about this product or the latest revision of this Instruction Sheet, visit the PASCO web site and enter
CI-6520A in the Search window.

Range*

Gain

Resolution

Accuracy

1

Calibration Factor

±1000 gauss

1X

0.5 gauss

10% of reading

100 gauss/volt

±100 gauss

10X

0.05 gauss

10% of reading

10 gauss/volt

±10 gauss

100X

0.05 gauss

10% of reading

1 gauss/volt

Address: PASCO scientific

10101 Foothills Blvd.
Roseville, CA 95747-7100

Phone:

916-786-3800 (worldwide)
800-772-8700 (U.S.)

Fax:

(916) 786-7565

Web:

www.pasco.com

Email:

[email protected]

1/4-20 threaded

hole

radial sensing

element

axial

sensing

element