P r o d u c t D e s c r i p t i o n

2

The Turbidity Sensor makes relative measurements of tur-
bid solutions with a particle size between 0 and 200
microns. Because turbidity is a time-dependent dynamics
phenomenon, factors such as particle density, particle size,
temperature, and pressure may cause reading variation,
especially in repeatability studies. For particle sizes above
200 microns, extract the large particles or allow more time
for measurement.

Suggested Activities

• Compare the turbidity of water sample from various

locations.

• Determine the rate of settling of a sample.

• Measure the formation of a precipitate.

Quick Start

• Plug the Turbidity Sensor into one of the PASPORT

input ports of a PASCO interface (such as the 850 Uni-
versal Interface or SPARK Science Learning System).

• NOTE: If more distance is needed between the sensor

and the interface, connect the sensor to the PASPORT
Extension Cable (PS-2500) and then plug the cable into
the interface.

• Put 6 milliliters (mL) of a fluid sample in a cuvette and

cap it. NOTE: Avoid touching the glass with your fin-
gers. Wipe the glass with a non-abrasive cleaning tissue
and/or silicone oil.

• Put the sample cuvette into the Sample Compartment

and close the lid.

• Start the PASCO data acquisition software (such as

PASCO Capstone). Set up a data display in the software.

• Begin recording data.

SETUP: For more information on setting up the software
and recording data, refer to the User’s Guide and online
help for the data acquisition software.

Calibration

Preparation Procedures: Regular Use

Calibration of the Turbidity Sensor is required the first
time the sensor is used. It is also necessary when you are
measuring solutions with varying temperatures, or when
using different cuvettes. All calibrations are stored inside
the sensor in flash memory.

• To prepare for calibration of the Turbidity Sensor, gen-

tly invert the 100 NTU Standard cuvette (included) five
times. Clean the outside of the cuvette.

• Put 6 mL of deionized water into an empty cuvette and

tighten the cap. Clean the outside of the cuvette.

• Connect the sensor to a PASPORT-compatible interface.

• Put the cuvette with the deionized water into the sensor

Sample Compartment and close the lid.

Calibration

• Press and release the green Calibration Button on the

sensor.

• The light-emitting diode (LED) in the button will come

on.

• In a moment, the LED in the button will begin to blink.

After a few moments of blinking, press the green button
again to stop the blinking.

• Replace the first cuvette with the 100 NTU Standard

cuvette and close the lid

• Press and release the green button. After the LED

begins to blink, press the button again to stop the blink-
ing.

• To verify, start the PASCO data acquisition software, set

up a Digits display, and start recording data.

• The Digits display should show approximately 100

NTU ±1 NTU.

• Stop recording data.

Sensor

PASPORT

Input Port

Interface