3 performance - model 705 interface, 4 process conditions – Orion Instruments Eclipse 705 User Manual
Page 54
54
57-600 Eclipse
®
Guided Wave Radar Transmitter
Reference Conditions
Reflection from liquid of selected dielectric at +70° F (+20° C) with 72" probe
Linearity
<0.5 inch
Measured Error
Upper layer ±1 inch
Interface layer ±1 inch (clean distinct interface required)
Upper Layer Dielectric
1.4–5.0
Interface Layer Dielectric
>15
Resolution
±0.1 inch
Repeatability
<0.5 inch
Hysteresis
<0.5 inch
Response Time
<1 second
Warm-up Time
<5 seconds
Operating Temp. Range
-40° to +175° F (-40° to +80° C)
LCD Temp. Range
-5° to +160° F (-20° to +70° C)
Ambient Temp. Effect
Approximately ±0.02% of probe length/ ° C
Humidity
0-99%, non-condensing
Electromagnetic Compatibility
Meets CE requirements: EN 61326
3.6.3 Performance - Model 705 Interface
Model
Coaxial
(7xA, 7xG, 7xT, 7xR)
Twin Rod
(7xB)
HTHP Coaxial
(7xD)
HP Coaxial
(7xP)
Steam
(7xS)
Maximum
Process Temperature
7xA:
+300° F @ 400 psig
(+150° C @ 27 bar)
7xG, 7xT & 7xR:
+400° F @ 270 psig
(+200° C @ 18 bar)
+400° F @ 275 psig
(+200° C @ 19 bar)
+800° F @ 1500 psig
(+427° C @ 103 bar)
+400° F @ 5500 psig
(+200° C @ 380 bar)
+650° F @ 2400 psig
(+340° C @ 165 bar)
Maximum
Process Pressure
1000 psig @ +70° F
(70 bar @ +20° C)
1000 psig @ +70° F
(70 bar @ +20° C)
6250 psig @ +70° F
(430 bar @ +20° C)
6250 psig @ +70° F
(430 bar @ +20° C)
3000 psig @ +100° F
(207 bar @ +38° C)
Maximum Viscosity
500 cp (Standard)
1500 cp (Enlarged)
10,000 Caged Coaxial
1500 cp
500 cp (Standard)
1500 cp (Enlarged)
500 cp (Standard)
1500 cp (Enlarged)
500 cp
Dielectric Range
≥1.4
≥1.9
≥1.4
≥1.4
>10
Hermeticity
N/A
N/A
Helium leak rate <10
-8
cc/sec
@ 1 atmosphere vacuum
N/A
Model
Rigid
(7xF)
Rigid
(7xJ)
Flexible
(7x1)
Flexible
(7x2, 7x5)
Hygienic
(7xF-E)
Paint
(7xF-P)
Maximum
Process Temperature
+300° F @ 400 psig
(+150° C @ 27 bar)
+605° F @ 1600 psig
(+320° C @ 110 bar)
+300° F @ 400 psig
(+150° C @ 27 bar)
+150° F @ 50 psig
(+66° C @ 3.4 bar)
+300° F @ 75 psig
(+150° C @ 27 bar)
160° F
(71° C)
Maximum
Process Pressure
1000 psig @ +70° F
(70 bar @ +20° C)
3550 psig @ +70° F
(245 bar @ +20° C)
N/A
75 psig @ +300° F
Atmospheric
Maximum Viscosity
10,000
(consult factory if severe agitation/turbulence)
2000
Dielectric Range
≥1.9
Hermeticity
N/A
3.6.4 Process Conditions
Not for direct insertion into boilers.