1 to install a model 7xf rigid single rod probe, 2 to install a model 7x1 flexible single rod probe – Orion Instruments Eclipse 705 User Manual
Page 13
13
57-600 Eclipse
®
Guided Wave Radar Transmitter
2. Ratio of Diameter: Length (A:B) is 1:1 or greater; any
ratio <1:1 (e.g., a 2"
× 6" nozzle = 1:3) may require a
Blocking Distance and/or DIELECTRIC adjustment
(see Section 2.6.5.2 Measurement Type: Level and
Volume).
3. No pipe reducers (restrictions) are used.
• Probe is kept away from conductive objects to ensure proper
performance. See Probe Clearance Table below. A lower gain
(increase in DIELECTRIC setting) may be necessary to
ignore certain objects (see Section 2.6.5.4 Measurement
Type: Interface and Volume).
2.4.3.1 To install a Model 7xF rigid single rod probe:
Make sure the process connection is at least 2" NPT or a
flanged mounting.
Carefully place the probe into the vessel. Align the gasket
on flanged installations.
Align the probe process connection with the threaded or
flanged mounting on the vessel.
For threaded connections, tighten the hex nut of the probe
process connection. For flanged connections, tighten flange
bolts.
Probe can be stabilized by placing into a non-metallic cup
or bracket at the bottom of the probe. A TFE bottom
spacer (P/N 89-9114-001) is optional for mounting into
a metallic cup or bracket.
NOTE: If the transmitter is to be installed at a later time, do not remove
the protective cap from the probe. Do not use sealing com-
pound or TFE tape on probe connection to transmitter as this
connection is sealed by a Viton
®
O-ring.
2.4.3.2 To install a Model 7x1 flexible single rod probe:
Make sure the process connection is at least 2" NPT or a
flanged mounting.
Carefully place the probe into the vessel. Align the gasket
on flanged installations.
Distance
to Probe
Acceptable Objects
<6"
Continuous, smooth, parallel conductive
surface, for example a metal tank wall;
important that probe does not touch wall
>6"
<1" (25mm) diameter pipe and beams,
ladder rungs
>12"
<3" (75mm) diameter pipe and beams,
concrete walls
>18"
All remaining objects
PROBE CLEARANCE TABLE
A
B
➄
➀
➁
➂
➃