beautypg.com

Health physics division, Nuclear characteristics, Functional features – Mirion Technologies SpiR-ID User Manual

Page 2: Identpro/sia identification algorithm, Accessories, Options

background image

Health Physics

Division

Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.

Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179

Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506
FIN-20101 Turku
Finland

T +358 2 468 4600
F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49
DE-22761 Hamburg
Germany

T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924

SpiR-ID

Detection & Identification - Risks Prevention

NUCLEAR CHARACTERISTICS

Detectors

SpiR-ID NaI: 3” x 1.5”

SpiR-ID LaBr: 1.5” x 1.5” LaBr

2 moderated LiI(Eu) detectors

GM tube for high gamma dose rate

Energy range

25 keV to 3 MeV (gamma)

0,025 eV to 15MeV (neutrons)

Measurement range

<0.01 µSv/h to 10 mSv/h ( 1µR/hr to
1 R/hr)

Identification

fast digital MCA 1024 channels

throughput >100 000 cps

range <0.01 to >25 µSv/h (137Cs)

detection and ID performance
overpasses ANSI N42-34, IEC62327,
IAEA NSS1 recommendations

Continuous spectra acquisition and
stabilization

no need of field calibration

Continuous count rate analysis
versus dynamic background

True dose rate calculation by spectra
weighting

Various spectra accumulation mode:

sliding / automated / manual

FUNCTIONAL FEATURES

Large TFT 3.5” VGA (640 x 480) sun

readable display

trend curve screen with level and ID

spectra and ID combined screen

event list screen with easy scroll

Alarm indication

sound or voice alarm and chirp

vibrator in the handle

colored LEDs

progressive alarm based on

identification and level

easy acknowledge

safety alarm

Routine mode / Expert mode

Embedded GPS

Data storage: SD card

Data retrieve and remote operation

USB link for PC host

External alpha/beta probe connection

IDENTPRO/SIA IDENTIFICATION

ALGORITHM

Designed for reliable decision with

short spectra

Solves mixed isotopes (up to 8)

Processes masking scenarios

Tolerant to shielding

Category, isotope, confidence level

indication

ENVIRONMENTAL, ELECTRICAL,
MECHANICAL CHARACTERISTICS

Temperature range: –20°C to 50°C

(-4°F to 122°F); -40°C on request

Humidity: up to 100%

Shock, vibration and drop resistant

MIL461D EMI compliant

Water and dust ingress IP65

Electrical:

Li-ion rechargeable, built-in charger

autonomy: typical 10 hours

accepts primary batteries for

immediate availability

Dimensions: 320 X 145 X 175 mm

3

(12.59 x 5.70 x 6.88 cu in)

Weight: NaI version 4.9 kg (10.8 lb),

LaBr version 4.02 kg (8.86 lb)

ACCESSORIES

Carrying case, carrying strap

External main power supply

SPIR-Companion software

OPTIONS

Alpha – beta probe

Embedded GSM / GPRS / Edge / 3G

Desk and wall holder

www.mirion.com

145203EN-D

SPIR Companion software

Common to all libraries

Source search library

Nuclear accident library (*)

(*) Firmware version V1.2.3 and up

40K, 226Ra + daughters, 232Th + daughters

NORM

SNM

merged

SNM

expert
Other

Medical

Industrial

Nuclear

U, Pu, 237Np
U, LEU, HEU, 233U, HBPu, MBPu, LBPu,
238Pu, 241Pu, 237Np
Neutron, Bremsstrahlung (ie Beta), Unknown

18F,/511keV, 51Cr, 67Ga, 99mTc, 99Mo,
103Pd, 111In, 123I, 125I, 131I, 133Xe,
153Sm, 201Tl
22Na, 54Mn, 57Co, 60Co, 75Se, 133Ba,
137Cs, 152Eu, 154Eu, 166Ho, 207Bi, 228Th,
241Am

18F/511keV, 60Co, 106Rh, 131l, 132l, 133l,
Xe133, 134Cs, 137Cs, 152Eu, 154Eu