Health physics division, Nuclear characteristics, Functional features – Mirion Technologies SpiR-ID User Manual
Page 2: Identpro/sia identification algorithm, Accessories, Options
![background image](/manuals/338025/2/background.png)
Health Physics
Division
Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
SpiR-ID
Detection & Identification - Risks Prevention
NUCLEAR CHARACTERISTICS
●
Detectors
○
SpiR-ID NaI: 3” x 1.5”
○
SpiR-ID LaBr: 1.5” x 1.5” LaBr
○
2 moderated LiI(Eu) detectors
○
GM tube for high gamma dose rate
●
Energy range
○
25 keV to 3 MeV (gamma)
○
0,025 eV to 15MeV (neutrons)
●
Measurement range
○
<0.01 µSv/h to 10 mSv/h ( 1µR/hr to
1 R/hr)
●
Identification
○
fast digital MCA 1024 channels
○
throughput >100 000 cps
○
range <0.01 to >25 µSv/h (137Cs)
○
detection and ID performance
overpasses ANSI N42-34, IEC62327,
IAEA NSS1 recommendations
●
Continuous spectra acquisition and
stabilization
○
no need of field calibration
●
Continuous count rate analysis
versus dynamic background
●
True dose rate calculation by spectra
weighting
●
Various spectra accumulation mode:
○
sliding / automated / manual
FUNCTIONAL FEATURES
●
Large TFT 3.5” VGA (640 x 480) sun
readable display
○
trend curve screen with level and ID
○
spectra and ID combined screen
○
event list screen with easy scroll
●
Alarm indication
○
sound or voice alarm and chirp
○
vibrator in the handle
○
colored LEDs
○
progressive alarm based on
identification and level
○
easy acknowledge
○
safety alarm
●
Routine mode / Expert mode
●
Embedded GPS
●
Data storage: SD card
●
Data retrieve and remote operation
○
USB link for PC host
●
External alpha/beta probe connection
IDENTPRO/SIA IDENTIFICATION
ALGORITHM
●
Designed for reliable decision with
short spectra
●
Solves mixed isotopes (up to 8)
●
Processes masking scenarios
●
Tolerant to shielding
●
Category, isotope, confidence level
indication
●
ENVIRONMENTAL, ELECTRICAL,
MECHANICAL CHARACTERISTICS
●
Temperature range: –20°C to 50°C
(-4°F to 122°F); -40°C on request
●
Humidity: up to 100%
●
Shock, vibration and drop resistant
●
MIL461D EMI compliant
●
Water and dust ingress IP65
●
Electrical:
○
Li-ion rechargeable, built-in charger
○
autonomy: typical 10 hours
○
accepts primary batteries for
immediate availability
●
Dimensions: 320 X 145 X 175 mm
3
(12.59 x 5.70 x 6.88 cu in)
●
Weight: NaI version 4.9 kg (10.8 lb),
LaBr version 4.02 kg (8.86 lb)
ACCESSORIES
●
Carrying case, carrying strap
●
External main power supply
●
SPIR-Companion software
OPTIONS
●
Alpha – beta probe
●
Embedded GSM / GPRS / Edge / 3G
●
Desk and wall holder
www.mirion.com
145203EN-D
SPIR Companion software
Common to all libraries
Source search library
Nuclear accident library (*)
(*) Firmware version V1.2.3 and up
40K, 226Ra + daughters, 232Th + daughters
NORM
SNM
merged
SNM
expert
Other
Medical
Industrial
Nuclear
U, Pu, 237Np
U, LEU, HEU, 233U, HBPu, MBPu, LBPu,
238Pu, 241Pu, 237Np
Neutron, Bremsstrahlung (ie Beta), Unknown
18F,/511keV, 51Cr, 67Ga, 99mTc, 99Mo,
103Pd, 111In, 123I, 125I, 131I, 133Xe,
153Sm, 201Tl
22Na, 54Mn, 57Co, 60Co, 75Se, 133Ba,
137Cs, 152Eu, 154Eu, 166Ho, 207Bi, 228Th,
241Am
18F/511keV, 60Co, 106Rh, 131l, 132l, 133l,
Xe133, 134Cs, 137Cs, 152Eu, 154Eu