Range limits and indication – Meriam M2110L SMART LEVEL GAUGE User Manual
Page 31
9R143 Dated 3/2011
29
RANGE LIMITS AND INDICATION
Sensor Hard
Soft
Certified
Soft
Hard
Type
Under Under
Range
Over
Over
Range Range
(F.S.)
Range
Range
DN
Under
-20%
1
-20 to 0%
0 to 100%
100 to
120%
Over
120%
D
I
Under
-20%
2
-20 to 0%
0 to 100%
100 to
120%
Over
120%
G
I
Under
-20%
3
-20
3
to
0%
0 to 100%
100 to
120%
Over
120%
C
I
NA
4
NA
4
0 to 100%
100 to
120%
Over
120%
A
I
NA
5
NA
5
0 to 100%
100 to
120%
Over
120%
Notes:
1
DN units will measure to -20% (or -10 PSID, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
2
D
I
units will measure to -20% (or -150 PSI, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
3
G
I
units will measure to -20% (or -10 PSIG, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
4
C
I
units will measure to negative barometric pressure (typically -14.7
PSIG) and to +120% of full scale pressure. Calibration certification is for
-14.5 PSIG to F.S. (Full Scale Only).
5
A
I
units will measure to +120% of full scale pressure. Calibration
certification is for 0 PSIA to F.S. (Full Scale Only).
Hard Over & Under:
Display will read "OP" (Under -20% or Over 120%).
Soft Over & Under:
Display will still read out at (± 20%).