Problem Summary
Resolution
Notes
(1x) Sanitize with 0.1N NaOH
(reservoir C)
(2x) Clean with 0.1N NaOH
(reservoir C)
(2x) Rinse (reservoir D)
(1x) Alcohol Flush with70%
isopropanol or ethanol alcohol
(reservoir B)
(1x) Prime
(4x) Rinse (reservoir D)
CAL (strip well A)
Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
(3x) Rinse (reservoir D)
VER (strip well B)
Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
(1x) Rinse (reservoir D)
Fluidics 1 (strip well C)
Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
Fluidics 2 (strip well D)
Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
(3x) Rinse (reservoir D)
Check the Cal/Ver Icon in the
System Monitor Bar located at
the bottom of the xPONENT
screen to confirm that the
operation has been completed
successfully.
If problem is not solved, contact Luminex Technical Support
1st Well “Low Bead Count”
Under the xPONENT 4.1 Admin
Tab. Verify that “Run Rinse
Before Batch Starts” is checked.
Rerun the plate .
Close and restart xPONENT.
Error at xPONENT Start
If you have configured your PC to
auto-start xPONENT when
booting, please disable this
function.
MAGPIX
®
User Quick Guide
20