Luminex xPONENT for MAGPIX IVD User Manual
Page 11

1. Open the Maintenance page, then the Cmds & Routines tab.
2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After
Storage routine performs the following commands:
• Prime
• Rinse
• Alcohol Flush
• Rinse
3. On the Cmds & Routines tab, select Prime from the Commands option section and
perform a Prime.
4. Add 70% isopropanol or ethanol to reservoir RB1 on the off-plate reagent block as
indicated on the Cmds & Routines tab.
NOTE: The rinse reservoir (RD1) should be empty.
5. Click Retract.
6. Click Run.
FIGURE 2.
Revive After Storage (Luminex) Routine
After the Revive After Storage routine is complete, run the System Initialization routine.
For In Vitro Diagnostic Use
Starting the MAGPIX System for the First Time
5