Magpix new system startup errors – Luminex xPONENT for MAGPIX IVD User Manual
Page 24

1. Adjusting the Sample Probe Height
2. Revive After Storage Routine (Luminex)
3. System Initialization
All system startup routines are run from the MAGPIX Off-Plate Reagent Block. For
detailed instructions on how to perform these operations, refer to the
MAGPIX System for the First Time
FIGURE 6.
MAGPIX Off-Plate Reagent Block
1
Reservoirs
2
Strip Wells
MAGPIX New System Startup Errors
Problem Summary
Resolution
Notes
Calibration Failure
(2x) Clean with 70% isopropanol
or ethanol alcohol (reservoir B)
(2x) Alcohol Flush with 70%
isopropanol or ethanol alcohol
(reservoir B)
(3x) Rinse (reservoir D)
(3x) Prime
Run Calibration Routine
Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
MAGPIX
®
User Quick Guide
18
For In Vitro Diagnostic Use