Create a protocol – Luminex xPONENT 4.2 for FLEXMAP 3D User Manual
Page 14
To run, print, and/or save a report:
1. Open the Results page.
2. Click the Reports tab.
3. In the Report drop-down list, select the report that you want to view.
4. In the Type drop-down list, select the type of report you want to view.
NOTE: Select dates if applicable.
5. Select the item for which you want to generate the report. If you are creating a Batch
Report, select the analytes to include in the report.
6. Click Generate.
7. Click Print to print the report, or Save to save the report as a PDF file.
Create a Protocol
You can create protocols with four types of analysis:
Quantitative - Analyzes unknown data generated with a standard curve.
Qualitative - Analyzes unknown data generated with cutoff ranges based on one standard.
Allele Call - Analyzes data generated as either heterozygotes or homozygotes, based on
specific cutoff ratios.
None - Used when analysis of the data will be carried out in another program. (xPONENT
®
reports are unavailable for data acquired with a protocol of type “none”.)
xPONENT
™
for FLEXMAP 3D
®
10