11. Hand-tighten the screw on the syringe arm.

12. Plug in the power cord and turn the Luminex 200 analyzer power

on.

13. Prime the system twice, watching for any leaks in the syringe

area.

14. When the prime finishes, close the center door.

Luminex 200 Analyzer
Ventilation Filter

‹

To clean the Luminex 200 analyzer ventilation filter:

1. Disconnect the Luminex 200 analyzer from ac power by turning

off the power switch on the rear of the analyzer, then unplug the
analyzer power cord from the wall source.

2. While facing the Luminex 200 analyzer, place your index finger

up under the right side of the analyzer (in the space between the
Luminex 200 analyzer and the Luminex XYP instrument). When
you feel the filter, push the filter toward the left of the analyzer.
See Figure 3-9.

3. Remove the filter from the left side of the Luminex 200 analyzer.

Figure 3-9 Luminex 200 Analyzer Ventilation Filter

4. Clean the filter with a vacuum or with distilled water. Stand the

filter upright to air dry.

5. Re-install it with the arrows facing up. The filter should click

into place.

6. Plug in and turn on the Luminex 200 analyzer power.