Required laboratory reagents, Luminex is 2.3 software, Luminex 200 performance specifications – Luminex 200 User Manual with IS 2.3 Software User Manual
Page 19: Speed -3 accuracy and precision -3

x
MAP Technology
The System
PN 89-00002-00-144 Rev. B
2 - 3
Required
Laboratory
Reagents
•
Household bleach
•
70% isopropanol or 70% ethanol
•
Mild detergent
•
Distilled water
Luminex IS 2.3
Software
Luminex IS 2.3 software provides complete control of the system
and performs data analysis. Your Luminex 200 system is preloaded
with the Luminex software. However, we supply a software CD
should you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.
Luminex 200
Performance
Specifications
Speed
•
Speed: greater than or equal to1.7 GHz Intel® Pentium® IV
processor with 256 MB RAM
•
USB communications link for fast data transfer
•
Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write CD
•
Installation: < 4 hours
•
System calibration: < 10 minutes
•
System controls: < 10 minutes
•
Barcode reader entry of sample IDs
•
Automatic post-analysis
•
Analyze one 96-well plate/hour depending on manufacturer’s kit
•
Up to 100 xMAP microsphere sets per sample
•
System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.
Accuracy and Precision
•
Sample uptake volume: ± 5%
•
Classification of xMAP microspheres: > 80%
•
Misclassification of xMAP microspheres:
≤ 2% - may vary by
xMAP microsphere product lines. Refer to the specific product
information sheet for further details.
•
Temperature control: 0°C to + 2°C of target