Luminex 200 System User Manual

x

MAP Technology

3 - 4

PN 89-00002-00-144 Rev. B

Figure 3-4 Refill Sheath Fluid Warning (Circled in Photo)

Refill the Sheath

Fluid Container

‹

To refill the sheath fluid container:

1. Release system pressure by removing the lid from the sheath

fluid container.

2. Refill the sheath fluid container.

If the sheath container ever runs dry, prime the system at least twice
until the air is removed from the system.

Empty the Waste

Container

‹

To empty the waste container:

1. Disconnect the waste container from the Luminex 200 analyzer.

2. Unscrew the waste container lid, being careful not to touch the

Gore-Tex

TM

seal. If the seal becomes wet or soiled, it may impair

venting.

3. Discard the waste from the waste container by appropriate

means.

4. Reconnect the waste container to the Luminex 200 analyzer and

replace the cap.

Any time the sheath container is disconnected from the Luminex
200 analyzer, you must remove air from the sample lines by
priming.

Note:

There is NO warning of

a high waste volume. Empty
the waste container each time
you fill the sheath container.