Install the luminex xyp instrument reservoir – Luminex 200 Installation Guide User Manual
Page 39

1. Remove the clear plastic shield that covers the sample probe area.
NOTE: Verify that the microtiter plate is not warped. Warped plates can lead
to incorrect probe height adjustment.
2. In a 96-well microtiter plate where overall height is no more than 0.75 inches (19 mm),
place the appropriate alignment tool in the plate:
NOTE: Alignment discs can be placed in any well as long as the well is
designated in the software.
For a standard plate with flat-bottom wells - stack two of the larger (5.08 mm diameter)
alignment discs together and place them into the selected well.
For a filter bottom plate - stack three of the larger (5.08 mm diameter) alignment discs
together and place them into the selected well.
For a half-volume plate with flat-bottom wells - stack two of the smaller (3.35 mm
diameter) alignment discs together and place them into the selected well.
For a round-bottom (U-bottom) plate - stack two of the smaller (3.35 mm diameter)
alignment discs in the selected well.
For a plate with conical wells - place one alignment sphere into the selected well.
3. Eject the plate holder. Place the 96-well microtiter plate on the Luminex XYP instrument
plate holder with position A1 in the top left corner.
4. Verify that the correct well location is selected in the Luminex software, and that you are
using the appropriate number of alignment discs. Retract the plate.
5. Loosen the front thumbscrew on the probe holder one-third to one-half turn. Pull it
upward until it touches the top of the adjustment slide. Tighten the thumbscrew.
6. Use the Luminex software to lower the sample probe.
7. Loosen the front thumbscrew. Gently push the probe down until it just touches the top of
the alignment discs or sphere.
8. Tighten the front thumbscrew.
9. Raise the sample probe using the Luminex software. Replace the plastic shield that
covers the sample probe area.
Install the Luminex XYP Instrument Reservoir
Use the Luminex XYP instrument reservoir for the Luminex 200 analyzer maintenance
functions.
To install the Luminex XYP instrument reservoir:
1. Eject the plate holder.
2. Insert the reservoir in the upper-right corner of the plate holder.
3. Retract the plate holder.
Luminex
®
200
™
Installation Manual
28
In Vitro Diagnostic Use