Sponsler, Inc.

T650N Series Cryogenic Truck Totalizer

pg

9

DOC# MN-650


Use the “System Test” S9 (Injects 500 Hz signal).

Example #6:


Calibration Factor = .5279

Time = 60 seconds


Divider = 100

T.C. Factor = 1.0529

Substituting: Quantity Displayed =

(500 X 60 X .5279)

X (1.0529)

100

=

(15837)

X (1.0529)

100

= 158.37

X (1.0529)


= 166.7


The Temperature Correction Factor (T.C.F.) is referenced at the Normal Boiling Point (NBP). The example above
indicates that the product temperature is below the NBP in that the T.C.F. > 1. Likewise, if the product temperature is
above the NBP, the T.C.F. will be < 1.

Universal Formula:

Quantity Displayed =

Hz X Time in Seconds X CF

X (Temperature Correction Factor)

Divider

(

)