beautypg.com

Liquid Controls T650N SERIES User Manual

Page 11

background image

Sponsler, Inc.

T650N Series Cryogenic Truck Totalizer

pg

11 DOC# MN-650



System Setup and Installation

The Temperature Compensator has been calibrated for the RTD and tested in the factory. Once the unit is installed,
the product selector switch must be set to the correct position for the product being delivered.

The following table summarizes the various positions of the product selector switch and the corresponding product.

MODEL 820 OPTION ‘IC’ EPROM

POSITION

PRODUCT

LOW LIMIT

REF TEMP

ALARM TEMP

HI LIMIT

0

Oxygen

-200.00

-183.11

-168.00

-140.50

1

Nitrogen

-210.00

-195.930

-182.00

-155.30

2

Argon

-189.00

-186.046

-168.00

-140.50

3

CO

2

-40.050

-16.388

-6.30

-6.30

4

N

2

O

-51.087

-34.679

15.30

15.30

5

Calif. CO

2

-40.00

-25.817

-7.00

-7.00

6

MAPP

-17.871

15.384

50.50

15.50

7

LPG

-40.00

15.384

49.50

49.50

8

Calibrate &

switch test

9

Self test



The following table summarizes the option jumpers and their functions. NOTE: Only one jumper should be installed!

OPTION JUMPERS

Jumper

Description

A

(*do not install) Momentarily shorted in probe calibration mode starts the dip switch
verification test.

B

When this jumper is installed the unit will not totalize if the alarm LED is on.

C

When the alarm LED is on... If the jumper is installed the compensator reverts to the
‘alarm’ temperature and corresponding TCF for the product selected. If the ‘C’ jumper is
not installed, the compensator will set the correction factor for no compensation (TCF =
1.00) and will output the actual temperature without modification.

D

When jumper is installed the pump control time delays will be activated.



Calibration

To calibrate the unit, first set the product selector switch to position 8 (calibrate). Next replace the temperature probe
with a 100 ohm .1 resistor and observe the (-) and (+) LED’s on the PCB. If the (-) LED is on, turn the calibrate pot
clockwise until the LED goes off. If the (+) LED is on, turn the calibrate pot counterclockwise until the LED goes off.
Once both LED’s are off, the unit is calibrated properly and is ready for use.
NOTE: Be sure to set the product selector switch back to the correct position for the product being delivered