Laurel Electronics LAUREATE SERIES Serial Comunications User Manual
Page 24
![background image](/manuals/336106/24/background.png)
-23-
31
Serial Cnfg1
Bit 7 6 5 4
3 2 1 0 Continuous Output Data Rate
(continued)
0 1 1 1
18.1
21.8
1 0 0 0
36.3
43.5
1 0 0 1
72.3
86.7
0 0 0
300 baud
0 0 1
600 baud
0 1 0
1200 baud
0 1 1
2400 baud
1 0 0
4800 baud
1 0 1
9600 baud
1 1 0
19200 baud
1 = Send filtered value (0= send unfiltered value)
2F
Filter
Bit 7 6 5 4
3 2 1 0
0 0 0 0
Auto Filter
0 0 0 1
Batch (16 samples) filter
Time constant
60 Hz 50 Hz
0 0 1 0
Moving Average
.07s .085s
0 0 1 1
Moving Average
.14
.17
0 1 0 0
Moving Average
.28
.34
0 1 0 1
Moving Average
.57
.68
0 1 1 0
Moving Average 1.13
1.36
0 1 1 1
Moving Average 2.27
2.72
1 0 0 0
Moving Average 4.53
5.44
1 0 0 1
Moving Average 9.06 10.88
1 0 1 0
Unfiltered
1 = High adaptive threshold(0 = low threshold)
1 = Display filtered signal(0 = display batch)
1 = Take peak of filtered signal(0 = unfiltered)
1 = Alarm from filtered signal(0 = unfiltered)
2E
Setup
Bit 7 6 5 4
3 2 1 0
EXT IN A
EXT IN B
0 0 0
Reset
Meter Hold
0 0 1
Reset
Peak Display
* Both ExtinA &
* 0 1 0
Meter Hold
Peak Display
ExtinB =
* 0 1 1
Meter Hold
Tare
Function reset * 1 0 0
Peak Display
Tare
1 0 1
Tare
Reset
1 1 0
External Decimal Point 1
1 1 1
External Decimal Point 2
1 = Coords of 2 points for scale,offset
0 = Scale and Offset direct parameters
0
Normal Format, Ser Com Continuous mode
1
Special Start & Stop characters
1 = 50 Hz DPM power (0 = 60 Hz power)
0 0
4 1/2 digit display, .1 deg for temp
0 1
Remote display
1 0
4 1/2 digits count by 10, .01 deg for RTD temp
1 1
3 1/2 digit display, 1 deg for temp