Temperature probe/ fixed at 25˚c/entered, Temperature sensor id, Test amount – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 220: The sequence is empty, Temperature probe/ fixed at 25°c/entered, Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 220
Temperature
Probe/ Fixed at
25°C/Entered
Probe
The sample or standard temperature can be measured using a tem-
perature sensor connected to the Temp socket of the Titration Man-
ager. Select a temperature sensor in the Method parameters menu.
Fixed at 25°C
If measurements are to be performed at a constant temperature of
25°C.
Entered
If the temperature is to be entered manually.
Select in:
Edit method menu
Edit reagent menu if Titre = Calibrate
Edit electrode menu if Calibrate request = Yes
Temperature
sensor ID
Name of the temperature sensor or name of the electrode with a built-
in temperature sensor.
Enter in:
Edit Electrode > Calibration parameters menu
Edit Method > Method parameters
Edit reagent > Reagent parameters
Range available:
16 characters
Test amount
Sample quantity to be placed into the beaker.
Enter in:
Menu Beaker (Menu Sequence/Sample stack then press 1 Sample
stack).
Refer to "Sample stack", page 202
.
Range available:
0,001 to 100000 (unit = Sample Unit)
The sequence is
empty
You are trying to run or complete a sample stack/sequence experi-
ment which contains no programmed methods.
Go to Sequence/Sample stack, Edit sequence screen and check the
sequence.