Minimum value - maximum value, Missing beaker (sac error), Titra lab – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 139

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 139
Minimum value -
Maximum value
If the acceptance criteria option has been set to Yes, enter the accept-
ance range for the result. If the result lies outside these limits, a "High"
or "Low" warning message appears and the result is rejected by the
instrument.
Enter in:
Edit method > Results menu (Acceptance criteria = Yes)
Edit reagent > Results menu (Acceptance criteria = Yes)
Edit method > QC data menu
Range available:
Minimum value = 0.0 to Maximum value
Maximum value = Minimum value to 10
10
Unit:
Sample method: Result unit
Reagent calibration: depends on the Reagent unit selected
(eq/l if N, meq/l if mN, mol/l if M, mmol/l if mM or mg/ml if mg/ml).
Missing beaker
(SAC error)
Beaker not detected. Solve the problem and restart the sequence
from the beaker it stopped (key 1 Resume analysis).
Warning!
Do not change the turntable before restarting the sequence from the
beaker it stopped (key 1) or before restarting the sequence from the
next beaker (key 2). The sample changer identifies a turntable only
when a new sequence is initialized (equivalent to a keystroke on 3
End of sequence followed by 1 Run sequence).
Causes:
1.
There is no beaker at the dedicated position and/or there is not
enough liquid in the beaker and the beaker cannot be detected.
Refer to "Beaker detection minimum height", page 66
Add solution in the beaker or desactivate the beaker automatic
detection by clearing the option "Beaker detection" in
TitraMaster 85 (refer to TitraMaster 85 on-line help, topics "Edit-
ing an application - Configuration").
2.
Using a level, check that the sample changer is installed on a flat
and horizontal bench surface.
3.
If points 1 and 2 have failed, adjust the beaker detector position
(refer to the User’s Guide of the sample changer, chapter 6
"Maintenance").