Ref. electrode conflict, Reject a result, Remove burette – Hach-Lange TITRALAB 860_865 Reference Manual User Manual
Page 176: Remove method from a sequence, Remove reagent, See "remove, Titra lab

Titra
Lab
®
TitraLab 860 and 865 Reference Manual
Page 176
Ref. electrode
conflict
Two reference electrodes are being used in the same beaker.
Change one of the reference electrodes in the electrode system.
Reject a result
Refer to "Result accepted (Yes/No)", page 184
.
Remove burette
Before removing, rinse the burette to remove the reagent.
1.
Enter the Burette functions menu:
2.
Press 6 Remove burette. The piston is moved to its lowest posi-
tion (filling operation).
3.
Check the burette data and press 1.
4.
Remove the burette and press
✓,
see "Install burette", page 125
Remove method
from a
sequence
1.
Select the sequence to be emptied,
.
2.
Press 2 Sequence/Sample stack then 3 Edit sequence.
3.
Select the method to be removed using the LEFT/RIGHT arrow
keys.
4.
Press 3 Delete.
5.
Press 1 Remove method then
✓ to confirm or press Esc to leave
the screen without removing the method.
The method is removed from the sequence but is not deleted.
To remove all the methods from a sequence,
.
Remove reagent
Use this procedure when you no longer wish to use the reagent. To
remove a reagent, run the rinse procedure
in the Burette functions
menu.
Refer to "Burette functions menu", page 69
.