Installation

3.5 Connection to process measuring instruments

The filtered sample is pumped through the sample suction tube (in
the sample tube) to the connected process measuring instruments.

The angle bracket enables the sample tube to be attached to a wall
at a distance of approx. 20 cm [7.87 in.] from the process
measuring instrument.

Table 1 Terminal assignment

Terminal number

Connection

1

N (230 V AC / 50–60 Hz, Option: 115 V AC)

2

L (230 V AC / 50–60 Hz, Option: 115 V AC)

7

N (sample tube heater)

8

N (suction tube heater)

9

L (sample tube heater)

10

L (suction tube heater)

17/18

Protective conductor for signal line screening

19

Protective conductor for suction tube heater

20

Protective conductor for sample tube heater

21/22

Potential-free fault contact (NO)

23/24

Potential-free warning contact (NO)