Malfunctions, causes, resolution
6.1 Error messages
Possible sensor error messages are displayed by the sc controller.
6.2 Warning messages
Possible sensor warning messages are displayed by the sc controller.
Table 1 Error messages
Error displayed
Definition
Resolution
MEAS OVERRANGE
Measurement range exceeded,
signals too small, probe can no longer
measure this concentration.
If error occurs more frequently: find another installation
location.
CAL. INSUFF. –
Calibration insufficient.
Probe requires another calibration point in a lower
concentration.
CAL. INSUFF. +
Calibration insufficient.
Probe requires another calibration point in a higher
concentration.
ZERO POINT
Calibration is
too close to the zero point.
Calibrate again with higher concentration.
CAL REQUIRED
No existing calibration
Calibrate probe.
EE RSRVD ERR
Error in the probe electronics
Call customer service department.
ERROR PROBE
Error in the probe electronics
Call customer service department.
LED FAILURE
Faulty LED
Call customer service department.
Table 2 Warnings
Warning displayed
Definition
Resolution
TEST / MAINT
Counter at zero
Call customer service department.