Operation, 2 sensor setup – Hach-Lange TSS EX1 sc User Manual
Page 17

17
Operation
17
Operation
17
Operation
OUTPUT MODE
Behavior of the outputs during calibration or while the zero point is being set
HOLD
ACTIVE
TRANSFER
SELECTION
SENS MEASURE
Current, uncorrected measurement value
FACTOR
Can be set from 0.10 to 10.00, a detailed description is provided in section
CALIBRATE
MEMORY
POINT 1
Calibration point 1 is recorded
POINT 2
Calibration point 2 is recorded
POINT 3
Calibration point 3 is recorded
CLEAR MEMORY
Clears the recorded values for all points.
POINT 1
Current calibration for point 1
POINT 2
Current calibration for point 2
POINT 3
Current calibration for point 3
SET CAL DEFLT
Security prompt, all calibration points are cleared
CONFIGURE
EDIT NAME
Name can include up to 16 characters,
FACTORY CONFIG: Device number
MEAS UNITS
TRB: (FNU, EBC, TE/F, NTU, FTU)
TS: (mg/L, g/L, ppm, %)
FACTORY CONFIG: FNU
PARAMETERS
TRB, TS, FACTORY CONFIG: TRB
RESPONSE TIME
1 to 300 s, FACTORY CONFIG: 60 s
LOGGER INTERVAL
10 s, 30 s, 1 min, 2 min, 3 min, 4 min, 5 min, 6 min, 10 min, 15 min, 30 min, FACTORY
CONFIG: 10 min
FACTORY CONFIG
Security prompt, reset to factory configuration for all menu items listed above.
TEST / MAINT
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)