4 menu structure, 1 sensor status, 2 sensor setup – Hach-Lange TSS EX1 sc User Manual
Page 16: 1 sensor status 4.4.2 sensor setup, Operation

16
Operation
16
Operation
16
Operation
4.4 Menu structure
Note: Refer to
Section 6 Malfunctions, causes, resolution
for a list of all error and warning messages
as well as a description of all necessary corrective actions.
4.4.1 SENSOR STATUS
SELECT SENSOR (if there is more than one sensor)
ERRORS
Possible error messages: MEAS. OVERRANGE, CAL. INSUFF. +/–, ZERO, CAL REQUIRED,
EE RSRVD ERR, ERROR PROBE, LED FAILURE
WARNINGS
Possible warning messages: TEST/MAINT
4.4.2 SENSOR setup
SELECT SENSOR (if there is more than one sensor)
WIPE
Triggers a wiping operation
CALIBRATE (turbidity)
OUTPUT MODE
Behavior of the outputs during calibration or while the zero point is being set
HOLD
ACTIVE
TRANSFER
SELECTION
SENS MEASURE
Current, uncorrected measurement value
FACTOR
Can be set from 0.10 to 10.00, a detailed description is provided in section
OFFSET
Can be set from –100 to +100, a detailed description is provided in section
CALIBRATE
MEMORY
POINT 1
Calibration point 1 is recorded
POINT 2
Calibration point 2 is recorded
POINT 3
Calibration point 3 is recorded
CLEAR MEMORY
Clears the recorded values for all points.
POINT 1
Current calibration for point 1
POINT 2
Current calibration for point 2
POINT 3
Current calibration for point 3
SET CAL DEFLT
Security prompt, reset to default calibration
CALIBRATE (TS content)