Communication menu – Hach-Lange ORBISPHERE 410 User Manual User Manual
Page 70
68
Communication Menu
Gas sensor calibration report example
Calibration report nb 1
Type . . . . . . . . . . . . . . . . . . . . .Sensor calibrated on this instrument
Date (yy.mm.dd - hh:mm) . . . . .05.02.17 - 18:40
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Calibration coefficient . . . . . . . .122.40 µA/bar
Ratio with ideal membrane . . . .92.72 %
Ratio with last calibration . . . . .97.35 %
Stability . . . . . . . . . . . . . . . . . . .0.000 %
Calibration mode . . . . . . . . . . . .Direct value
Calibration value . . . . . . . . . . . .0.750000
Gas unit. . . . . . . . . . . . . . . . . . .bar
Gas phase. . . . . . . . . . . . . . . . .Liquid
Liquid. . . . . . . . . . . . . . . . . . . . .Water
Interferences . . . . . . . . . . . . . . .Disable
Temperature . . . . . . . . . . . . . . .20.1 °C
Barometric pressure . . . . . . . . .1.020 bar
External pressure . . . . . . . . . . .0.976 bar
Sensor current. . . . . . . . . . . . . .79.857 µA
Smart gas sensor connection calibration report example
Calibration report nb 1
Type . . . . . . . . . . . . . . . . . . . . .Sensor calibrated on another instrument
Date (yy.mm.dd - hh:mm) . . . . .05.01.17 - 15:30
Operator . . . . . . . . . . . . . . . . . .Armstrong
Calibration coefficient . . . . . . . .62.75 µA/bar
Connection date . . . . . . . . . . . .05.03.15 - 12:26
Serial number . . . . . . . . . . . . . .1
Membrane . . . . . . . . . . . . . . . . .2956A
Barometric sensor calibration report example
Calibration report nb 1
Date (yy.mm.dd - hh:mm) . . . . .05.02.16 - 20:38
Operator . . . . . . . . . . . . . . . . . .jp
Operator ID . . . . . . . . . . . . . . . .3
Former barometric pressure . . .0.970 bar
New Barometric pressure . . . . .0.971 bar
User action log file example
The "User action log file" below contains 3 user actions.
Nr
mm/dd hh:mm:ss User
ID
User Name
Action ID
Description
1 1/21
15:13:44
1007
Armstrong
140
Measurement config.
0 1/21
15:13:27
1007
Armstrong
132
Identification
2 1/21
15: 9:15
1007
Armstrong
132
Identification
Measurement file example
6 measurements are described below:
Nr mm/dd hh:mm:ss
Gas
Temp
Mask
Fluor. phi Barom Ext P. Index
[ppb]
[°C]
[°] [bar] [bar]
0
2/17
21:15:37
75.051
20.1
400
26.039 1.005 0.000 2271
1
2/17
21:15:27
75.043
20.1
400
26.045 1.005 0.000 2266
2
2/17
21:15:17
75.047
20.1
400
26.052 1.005 0.000 2261
3
2/17
21:14:57
75.044
20.1
400
26.041 1.005 0.000 2256
4
2/17
21:14:47
75.047
20.1
400
26.038 1.005 0.000 2251
5
2/17
21:14:37
75.050
20.1
400
26.054 1.005 0.000 2246