4 measurement board, Installation 3.7.4 measurement board – Hach-Lange ORBISPHERE 410 User Manual User Manual
Page 32
30
Installation
3.7.4 Measurement board
The different measurement boards for the EC and TC sensors are illustrated in
and
below. The type of board is easily identified by the color of the J8 connector. For EC
boards this connector is colored orange, and for TC boards it is colored black.
The colors indicated are the wire colors in the sensor cable.
NOTICE
It is extremely important that sensors are connected to the correct measurement board.
Connecting a TC sensor to an EC measurement board (and vice versa) will cause irreparable
damage to the measurement board.
Figure 14 EC Measurement board
Figure 15 TC Measurement board
Figure 16 Connector J7
Figure 17 Connector J8
Connector J7 (inputs & outputs)
Measurement alarms relays
1.
Common
2.
Output relay 1
3.
Output relay 2
4.
Output relay 3
Note:
Relays N.O. or N.C. depends on the jumper
Analog current (or voltage) outputs
5.
GND
6.
Output 1
7.
Output 2
8.
Output 3
Digital inputs
9.
EC sensor: Not used
TC sensor: Hold input (short circuit to pin
12)
10.
Not used
11.
Not used