3. Installation
3.3. Cell cables
Compensation of the cable length
To minimize the effect of the cable resistance in a conductivity measurement, the cable length is compensa-
ted by the 4-wire system. With the Pt 1000 temperature sensor, the cable resistance has little importance.
However, in order to reduce the resistance of the conductor to a half, 2 cables have been connected to
the ends of the sensor.
3.3.1. Cells with MP-5 connector
Notas importantes:
• El si628 C does not require ground connec-
tion (terminal 1).
• Connect the ground only when there could
be interference problems.
DANGER
Relays:
- Limit the current up to 3 A by using
one fuse.
-
Do not connect cables with voltage
higher than 24 V.
Terminal assignment (rear panel)
Transmitter si628 C
RED
VCA
CELL
4-20 mA
L 1
L 2
S/N:
CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona
Made in Spain
IP 54
TERMINALS
COLOUR
si628 C
(wall mount)
si628 C
(panel mount)
6
16
White
6
16
Brown
7
15
Pink
7
15
Yellow
8
14
Shield
9
13
Green
10
12
Blue
11
11
Gray
12
10
Red
MP-5 connector, 5 contacts