Specifications

Specifications are subject to change without notice.

Specification

Details

Measuring range

0 to 10,000 ppb freely programmable
0 to 200 ppm with K-Kit option

Accuracy

Non-cationic application: ± 0.1 ppb or ± 5% of reading, whichever is
greater
Cationic application: ± 2 ppb or 5% reading, whichever is greater

Repeatability

< 0.02 ppb or 1.5% reading, whichever is greater within 10°C variation

Average response times at 25°C
with a maximum ΔT =15°C between
channels

T90% ≤ 10 min

Concentration step

from one channel to

another

Max. temp variation

(°C)

Time to reach accuracy

0.1 ppb or 5%

up

down

0.1 ↔ 5 ppb

3

9 min

27 min

0.1 ↔ 50 ppb

3

11 min

41 min

0.1 ↔ 200 ppb

3

9 min

45 min

< 0.1 ↔ 1 ppb

3

29 min

36 min

0.1 ↔ 50 ppb

15

11 min

41 min

Electrode type

pH glass electrode

Number of channels

1 to 4

Interference phosphate 10 ppm

Measurement variation less 0.1 ppb

Sample temperature interference

< 0.5% / °C

Typical environment

Power station / indoor / demineralized water plant or instrumentation room

Suspended solids

< 2 NTU, no oil, no grease

Temperature range for storage

-20 to 60°C (2 to 140°F)

Relative humidity

10 to 80%

Ambient temperature

5 to 50°C (41 to 122°F)

Sample temperature variation

Stabilization in 10 mins from 15°C to 30°C
Use the static heat exchanger system when the temperature difference
between samples is higher than 15°C

pH range of sample

Non-cationic application: 6 to 10 pH
Cationic application: 2 to 10 pH

Flow rate

6 to 9 L/hour

Pressure

0.2 to 6 bar (3 - 87 psi)

Acidity

Less than 250 ppm (equivalent CaCO

3

)

Power supply voltage fluctuation

± 10%

Over voltage category

2 (according to standard EN 61010-1)

Pollution degree

2 (according to standard CEI 664)

Altitude

< 2000 m