Cashco 764PD Pressure Controller User Manual
Page 6

IOM-764P / PD
6
7. Remove the three sensor O-rings (5) from the
sensor (2.1).
NOTE: Typically, additional dis-assembly of 
the sensor sub-assembly (2) is not required. 
If it is necessary to replace any of the sensor 
sub-assembly parts (2.1, 2.2, 2.3, 2.4, 2.5) a 
complete new sensor sub-assembly must be 
installed. Contact the factory for assistance.
8. Take all the metal parts except the gauge (11)
and nameplate (10) and clean in suitable solvent. 
Clean and air blow all drilled air passages.
9. Install pressure gauge (11) into the pneumatic
hous ing (1) using suitable thread sealant.
10. Install new O-rings (5) on sensor sub-assembly
(2). Check the two dimensions given in Figure 10 
of the sensor sub-assembly as “fi nal set tings”. 
2. Remove cap screws (19) and nuts (18). Re move
diaphragm upper case (13) and upper diaphragm 
O-ring (16). Remove diaphragm (12), and SST 
diaphragm cover (17) if sup plied. Remove lower 
case diaphragm O-ring (16) and pusher post O-ring 
(23) for 764PD. In spect for any corrosion damage 
and/or per ma nent deformation of the diaphragm 
(12) and diaphragm cover (17); if bent, replace.
3. For 764PD, separate at the screwed joint between
the diaphragm lower case (14) and the adapter (21) 
turning the diaphragm case (14) CCW (viewed from 
di a phragm case end). When separated, remove 
the adapter O-ring (22) and replace. Reinstall the 
adapter (21) and diaphragm lower case (14).
4. For the 764PD, replace diaphragm O-ring (16) for
the lower case (14) and the pusher post O-ring 
(23).
5. Replace diaphragm (12) and diaphragm cover(s)
(17), if supplied. Assure that the di a phragm cover 
(thin SST material) (17) is placed on side(s) ex-
posed to the process fl uid(s). (764PD requires both 
sides to be cov ered, if supplied with diaphragm 
covers (17).
6. Replace with new upper case diaphragm O-ring
(16), upper case (13), and cap screws with nuts 
(19 and 18).
7. Recalibrate per Section VI.
C. Pneumatic Circuit Overhaul:
1. Remove output pressure gauge (11) and test
cal i bra tion. If gauge is off ± 2 psi (±0.14 Bar) 
recommend replacement.
2. Remove lower diaphragm case (14) from the
pneu mat ic housing (1) at the point where screwed 
to geth er. Rotate the diaphragm case (14) CCW 
(viewed from di a phragm case end) to loosen. For 
764PD: Remove adapter (21) together with lower 
diaphragm casing (14) at the screwed joint be-
tween the pneumatic hous ing (1) and the adapter 
(21). Rotate the adapt er (21) CCW (viewed from 
diaphragm case end) to loosen. Once loos ened, 
ori ent the unit along a hor i zon tal axis to pre vent 
internal parts from fall ing from within when sepa-
rated. DO NOT RE MOVE PUSHER-POST (24) 
FROM 764PD.
3. Remove sensor plate (15.1) (15).
4. Remove nameplate screw (9), nameplate (10),
and four housing screws (8). Grasp the sensor
sub-assembly (2) end plate (2.2), and pull it 
outwards from the housing (1). Watch for the 
sensor pin (2.4) as the sensor sub-assembly 
(2) is withdrawn, as pin (2.4) easily falls out and 
could be lost.
Figure 9: Sensor Sub-Assembly (2)
5. Remove sensor plate O-ring (6).
6. Take note of the position of the PB needle
valve (3) by rotating it CW until it seats: count 
the exact amount of rotation in 1/8 rev. in cre-
 ments; record the count in Table 3. Re move 
the PB needle valve (3) with O-ring (7) by 
rotating CCW. Remove O-ring (7).
TABLE 3
Part
DESCRIPTION 
OF ROTATION
NO. OF
REVOLUTIONS
PB Needle 
Valve (3)
No. of Revolutions rotated CW to 
fully seat
