Out (output) – setting the output signal – Burkert Type 8070 User Manual
Page 21
19
SettingandFunctions
9.4.3.
Out (Output) – setting the output signal
In this parameter, define whether the device is to work as a flow transmitter or a pulse divider.
setting as flow transmitter (4 - 20 mA):
When changing the unit (e.g. from liters to gallons) the limit values for the current output are not converted
automatically
→
Set the lower and upper limit values for the flow rate range associated to the 4-20 mA output, in the unit set within the
UNIT parameter.
The lower limit value is marked by an L (low) on the display and the upper limit value by an H (high).
setting as pulse divider (PULS):
→
Set the volume, in the displayed unit, for each pulse emitted on the NPN and PNP outputs.
Setting in the parameter:
* Order of value input:
1. Select the decimal point place
(each time the key is pressed: one position to
the left)
2. Set numerical values
Status display changes to "active"
set the flow rate range *
set the volume pro pulse, in the displayed unit *
Lower limit value
Upper limit value
select
flow
transmitter
select
pulse divider
mA
l
L = low
H = high
Fig. 13: Out – Setting the output; operation as flow transmitter or pulse divider