Model 5861e – Brooks Instrument 5861E User Manual
Page 8
1-2
Section 1 Introduction
Installation and Operation Manual
X-TMF-5861E-MFM-eng
Part Number: 541B107AAG
November, 2008
Model 5861E
Control / Usable Range
50 to 1
Sensitivity to Mounting Attitude
±0.5% F.S. maximum deviation from specified accuracy after rezeroing under
200 psig
Temperature Sensitivity
Zero: Less than ±0.075% F.S. per degree C
Span: Less than ±1.0% F.S. shift from original calibration over 10-50°C range
Pressure Sensitivity
0.03% per PSI up to 200 PSIG
Power Supply Sensitivity
±0.09% full scale per % power supply voltage variation
Output Signal
0 to 5 Vdc into 3000 ohms (or greater) load. Maximum ripple 3 mV
Leak Integrity
1 x 10
-9
atmosphere scc/sec. Helium
Power Requirements
+15 Vdc (±5%) at 35 mA dc
-15 Vdc (±5%) at 35 mA dc
1.05 watts power consumption
Materials of Construction
Fittings and Transducer Assembly - Wetted parts 316 stainless steel
O-rings and Gaskets - Standard: Viton
®
fluoroelastomers and Buna-N;
Optional: Kalrez
®
Electrical Connections
D-connector, 15-pin type (DA-15P)
Dimensions
See Figure 2-1