Model 5861e – Brooks Instrument 5861E User Manual
Page 20
3-6
Model 5861E
Section 3 Operation
Installation and Operation Manual
X-TMF-5861E-MFM-eng
Part Number: 541B107AAG
November, 2008
3. Adjust the linearity potentiometer for an output equal to the new TP2
voltage and then repeat step k.
Note: The voltage at TP2 can range from -10 to +3 volts, however, it is
recommended that this voltage stay between -2.5 and +2.5 volts for proper
operation. If the recommended voltage range is exceeded the desired
accuracy and/or signal stability may not be achieved. If one of the limits is
reached, check the restrictor sizing. Refer to section 4-6.
1 to 5 Vdc: 1
Vdc
0 to 20 mA: 0 mA
4 to 20 mA: 0 mA
In the Safe State, the valve will be unpowered. This means that for
Normally Closed valves, the valve will stay closed and for Normally Open
valves, the valve will stay open.
3-5 Response (Flow Output Signal)
To achieve the proper response characteristics the response
compensation circuit must be adjusted. This adjustment is performed by
observing the output signal of the meter when the flow is suddenly
stopped. Place a metering valve upstream of the Model 5861E to control
the flow rate. A Brooks NRS
TM
precision metering valve is well suited for
this application. Also place a fast acting shut-off valve immediately
downstream of the flowmeter. A solenoid valve is ideal for this, but a
manual toggle valve will do. Keep the length of interconnecting tubing as
short as possible between the valves and the Model 5861E since the
tubing can have a dampening effect on the flow and the gas may not stop
flowing the instant the downstream valve is closed as desired. Adjustment
of the fast response circuit will not alter the steady state accuracy of the
flowmeter as adjusted in Section 3-4.
Note: This procedure requires an oscilloscope, chart recorder, or a DVM
with a sample speed three samples per second or greater to monitor the
rate of change of the output signal during the test. Monitor the output signal
at pin 2 of D-Connector. TP4 may be used for circuit common.
a. With the shut-off valve open adjust the metering valve so that the output
voltage of the Model 5861E is 4.050 to 5.000 Vdc. Allow the output to
become stable at this setting.
b. Close the shut off valve to stop the flow. Observe the output signal as it
decays.
c. The behavior of the output signal during the transition between 100%
and 0% flow indicates the adjustment required of the anticipate
potentiometer. Refer to Figure 3-4.
1. If the flow signal decays to -0.05 to -0.5V then rises to 0.0V, the
anticipate potentiometer is properly adjusted.
2. If the flow signal decays rapidly and goes below -0.5V before rising to
0.0V, the anticipate potentiometer must be adjusted clockwise and
steps a and b repeated.
3. If the flow signal decays slowly and does not go below -0.05V, the
anticipate potentiometer must be adjusted counterclockwise and
steps a and b repeated.