Brooks Instrument SLA5810/20 User Manual
Page 9

1-3
Models SLA5810 and SLA5820
Section 1 Introduction
Installation and Operation Manual
X-PR-SLA5800-PC-eng
Part Number: 541B091AAG
July, 2010
Leak Integrity
Inboard to Outboard: 1x10
-9
atm scc/sec Helium max.
Helium leak test performed without fittings
Maximum Fluid Operating Temperature
0°C to 65°C (32°F to 149°F)
Ambient Temperature Limits
Operating: 0°C to 60°C (32°F to 140°F)
Non-operating: -25°C to 100°C (-13°F to 212°F)
PHYSICAL:
Materials of Construction
Wetted parts - stainless steel
Viton
®
fluoroelastomers
Optional:
Buna-N
Kalrez
®
Teflon
®
/Kalrez
Outline Dimensions
Refer to Figures 1-3 to 1-5
Process Connections
Refer to Figures 1-3 to 1-5
ELECTRICAL CHARACTERISTICS:
Electrical Connections
Analog option: 15-pin D-Connector, male (See Figure 1-1)
Digital I/O:
DeviceNet: 5-pin Micro-Connector, male
F
OUNDATION
Fieldbus:
one 4-Pin M8 connector for power and
one 4-Pin M12 connector for communications
Power Supply Voltage
Analog option:
+15 Vdc, ±5% (traditional -15 Vdc pin is ignored) or
+24 Vdc ±5%
Digital I/O:
DeviceNet I/O: 11-25 Vdc
F
OUNDATION
Fieldbus I/O: 14-27 Vdc
Power Requirements:
Watts, typical Watts, max.
Analog I/O option, no valve:
1.6
1.8
Analog I/O option, with valve:
3.6
4.0
Digital I/O option, n.v.:
3.6
4.0
Digital I/O option, w.v.:
6.9
7.6