Bray 5000_6000 Series User Manual
Page 3

Installation and Maintenance Manual
Series 5000 and 6000 Ball Valves
Date: August 2011 / Page 3 of 6
®
A Subsidiary of BRAY INTERNATIONAL, Inc.
FLOW-TEK,
Inc.
Tel:
832.912.2300
© 2011 Flow-Tek, Inc.
8323 N. Eldridge Pkwy #100
Fax: 832.912.2301
Houston,
Texas
77041
www.flow-tek.com
closed. Lift the ball from body, using a strap and lift
device if necessary.
NOTE: Extreme caution should be taken to avoid
damage to the ball.
7. The stem must be removed from inside the body - a
tap on the top of the stem should loosen it. The thrust
washer should come out with the stem. Then remove
the stem packing.
Disassembly for servicing seats and stem seals may be
performed without removing the end flanges from the
pipeline.
This feature allows the removal of the valve from service
without disturbing piping connections.
5. VISUAL INSPECTION:
Clean and inspect all metal parts for damage and wear.
It is not necessary to replace the ball and stem unless
the sealing surfaces have been damaged by abrasion
or corrosion. Replacement of all soft parts is strongly
recommended whenever the valve is disassembled for
reconditioning. This is the surest protection against
leakage. The replacement parts can he ordered in kit form.
NOTE: The valve may be assembled and operated dry
where no lubricants are allowed in the system; however,
a light lubrication of mating parts will aid in assembly
and reduce initial operating torque. Any lubricant used
must be compatible with the intended line fluid.
6. RE-ASSEMBLY:
Refer to the illustrations on page 4 and 5.
1. Hold the body horizontally in a vise firmly enough
that it will not move but do not crush it.
2. Put thrust washer (12) on valve stem (5) and slide
the stem into the stem hole from inside the valve
body (1);
3. Slide the stem packing components (14) onto the stem
until they are bottomed in the stem hole;
4. Install the packing gland nut (15) into the body over
the stem and packing. Tighten finger tight.
5. Turn the stem until the upper stem flat is perpendicular
to the valve centerline (closed, so that the ball can
fit onto the stem):
6. Install the ball (3) inside the valve body. The lower
stem flats will fit into the slot in the ball;
7. Turn the stem so that the ball is in the open position
(the ball will not fall out of the valve);
8. Install the seats (4) and body seals into the body
ends. Make sure the spherical curvature side of the
seat will face the ball;
9. Lift the valve into position between the flanges in the
pipeline. Use caution that the seal components do
not shift position or fall out.
10. Install the end cap bolts (9), lock washers (7) and
nuts (8). Alternately cross tighten all nuts gradually
up to the values indicated in Table 1. Using a torque
wrench is recommended.
11. Install and secure the handle (25) with the handle
washer (19) and handle nut (26).
12. Cycle the valve slowly several times. Cycling the
valve slowly will insure that the seats will align with
and conform to the ball. A fast turning motion may
damage the seats before they have a chance to
form the proper seal:
Refer to the instructions given in section 2, above, to
re-commission the valve after installation.
For automated installations, follow the instructions
provided with actuation instrumentation to correctly
set up and recalibrate the installation.
Table 1 - Tightening torques for end cap nuts
VALVE SIZE
1/4” 3/8” 1/2” 3/4”
1”
1 -1/4” 1-1/2”
2”
2-1/2”
3”
4”
NutTorque (lb-in)
70
70
155
155
155
300
300
(Lb-ft)
25
25
43
58
83
83