beautypg.com

Bray 5000_6000 Series User Manual

Page 2

background image

Installation and Maintenance Manual

Series 5000 and 6000 Ball Valves

Date: August 2011 / Page 2 of 6

®

A Subsidiary of BRAY INTERNATIONAL, Inc.

FLOW-TEK,

Inc.

Tel:

832.912.2300

© 2011 Flow-Tek, Inc.

8323 N. Eldridge Pkwy #100

Fax: 832.912.2301

Houston,

Texas

77041

www.flow-tek.com

end being tightened. Tightening by using the valve

body, handle or the opposite end cap can seriously

damage the valve; Use a wrench on both the second

end cap and pipe to avoid applying torque to the

body through the bolting.

3. In some applications, threaded end valves may be

back-welded on site. Use caution. Do not overheat the

valve body. Confirm that the end cap bolt holes are

aligned with the body screw holes on both ends and

that the end caps are parallel and correctly spaced

to fit the valve body before re-assembly or back

welding. Make up both of the threaded connections

normally and then remove the body assembly as

described below for weld-end connections before

back welding the threaded connection.

4. Do not use the body screws to pull the pipe ends

together to correct misalignment.

INSTALLATION OF WELD-END VALVES

Insure that weld procedures are implemented that are

compatible with the materials used.

1. Tack weld the assembled valve to the pipe at four

points on both end caps.

2. With the valve in open position, (the valve is in the

open position when the handle is parallel to the axis

of the valve and pipe), loosen all body bolts, take

the body out.

3. Turn the handle in the half open position. Remove the

seats and body gaskets.

4. Turn handle in closed position and remove ball.
5. Place all removed parts in a clean and secure place.
6. Replace the body and all bolts and tighten them slightly.

This operation is very important, so that the body and

end caps remain perfectly parallel, thus preventing any

leakage at the body joint after welding.

7. Finish welding both end caps onto the pipe. Use

temperature indicating crayons to insure that the

body temperature doesn’t exceed 200°F in the area

of the neck or component/seal damage may occur.

8. When cool, remove the body, clean the end caps,

and reassemble per the instructions given below.

9. Place the bodies between the two end connections;

then replace all bolts and cross tighten.

10. (See table 1. Tightening torques for end cap screws).

4. DISASSEMBLY & CLEANING PROCEDURES:
CAUTION: Line must be depressurized before disassembly.

Valve should be cycled after draining the system to relieve

any pressure trapped in the valve body cavity. Ball valves

can trap pressurized media when closed. Flush line with

the valve 1/2 open to remove hazardous media. If the

valve has been used to control hazardous media, it must

be decontaminated before disassembly.

WARNING: DO NOT attempt to remove packing

components or any other valve parts while line is under

pressure!

CAUTION: Properly support the valve before removing

or loosening any body bolts. The larger sizes are heavy

and may cause injury if allowed to swing or fall freely.

Remove the actuation air and power connections and

confirm that there is no stored energy in the actuation such

as compressed springs or trapped air before beginning.

Stored energy devices can cause serious injury if the

energy is released without warning.

DISASSEMBLY FOR STEM AND SEAL SERVICE:

1. It is not necessary to remove the flanges from the

pipeline. Removing the actuation and controls may

make service easier.

2. Remove one of the upper body bolts and loosen the

remaining bolts. The center body section containing the

ball and seals will swing out for service, leaving the end

caps in place on the pipe ends. See illustration below.

3. To completely remove the valve from the pipeline,

remove the remaining upper body bolt and the body

will lift free. Gently support the body in a vise to

complete the remainder of the disassembly. To avoid

damage to the seal surfaces, hold the body horizontally

across its diameter, not across the end faces.

4. Remove the seats and body seals.

5. Remove the handle nut and handle. Remove the

gland nut.

6. To remove the ball, rotate the stem so the ball is fully