12. Press Esc key to exit.
CH01 CC OFF
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
13. Press Shift + ⑦ to enter System menu. Press key to select source> and press Enter. Select MANUAL 14. Press Esc key to exit. CH01 CC OFF 15. Press On/Off key to turn on the electronic load’s input. CH01 CC ON 16. Press Tran key on channel 1 module to enable transient operation. 17. Press Trig key on mainframe panel to trigger transient operation. CURRent:TRANsient:MODE CONTinous
EXTERNAL
HOLD
BUS
TIMER
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W
Transient Operation Programming Example
If in remote control mode, you can use the following example commands for transient operation
(refer to MDL Series Programming Guide for more information).
CURRent:TRANsient:ALEVel 10
CURRent:TRANsient:AWIDth 0.2ms
CURRent:TRANsent:BLEVel 2
CURRent:TRANsient:BWIDth 0.3ms
TRANsient ON
TRIGger:IMMediate