Toggle

In toggle mode, the electronic load will switch between A level and B level when receiving a trigger
signal after the transient operation is enabled. The following picture shows the current waveform
in toggle transient operation.

Figure 27 - Toggle Transient Operation

Setting Up A/B Transient Operation

The following is a short tutorial of how to set up A/B transient operation for your electronic load.
In this example, we will set up rise speed at 1 A/us and fall speed at 2 A/us. The electronic load will
be in continuous transient mode and switch between 10 A and 2 A in durations of 0.002 s and
0.003 s respectively.


Note : When setting up steps, make sure all transient level, slew, and width parameters are
within the modules’ specified limits.