Overpower protection (opp) – B&K Precision MDL Series - Manual User Manual
Page 13
13
Panel Operations
VFD Display
1. Power on the electronic load.
Self-test
2. Press Shift + ⑧ to enter Configuration menu.
Sync On Set
Von
Meter
Protect
List
Ext. Ctrl Set
About
Exit
3. Press key to select
menu.
Max Power Set
Alimit State
Alimit Point
Alimit Delay
Plimit Point
Plimit Delay
OnTimer State
OnTimer Set
Exit
4. Press key to select
and press Enter to confirm.
On
Off (Default)
5. Press key to select
current value and press Enter to confirm.
Current Limit
Set= 40.000A
6. Press key to select
time before alarm and press Enter to confirm.
Current Limit
Delay= 3S
7. Press Esc key to exit menus.
If the electronic load’s current value is above the set overcurrent protection value, the electronic
load will automatically turn off and the VFD will display OCP. At the same time, the OC and PS bits
in the status register will be set and remain until they are reset.
Operations to Clear the OCP State
Check whether the input current is within the electronic load’s rated current or the programmed
protection current ranges. If it is outside the range, disconnect the device under test. Then press
any key on the front panel or remotely send SCPI command PROTection:CLEar. The OCP displayed
on the front panel will turn off and the load exits OCP protection state.