Measuring, Compression joint, Digital milliohm meter – B&K Precision 310 - Manual User Manual

Page 9

background image

MEASURING

• Perform the preliminary checks before proceeding with

measurement and ensure that the precautions listed are
observed.

• Connect the test leads (color coded) to the instrument as shown.
• The current test leads must always be outside of the potential

test leads.

• Please note that the shorter the potential test leads, the better

long potential test leads will pick up noise.

• Screened test leads are recommended for better environmental

noise rejection.

-7-

R

X

R

B

R

A

Compression

Joint

DIGITAL MILLIOHM METER

Ÿ Short Test duration if TEST
is depressed for less than
2Sec.

Ÿ Long Test Duration if TEST
is depressed for more than
3Sec.