SPECIFICATIONS

ELECTRICAL
Measuring Ranges

0-200.0m ohms in steps of 100µ ohm
0-2000m ohms in steps of 1m ohm
0-20.00 ohms in steps of 10m ohm
0-200.0 ohms in steps of 100m ohm
0-2000 ohms in steps of 1 ohm

Accuracy

±0.5% of reading ±2 digits over the
Operating temperature range, -15°C to
+55°C, with the supplied test leads.

Test Current

1mA => 2000 ohms range.
10mA => 200 / 20 ohms ranges.
100mA => 2000m / 200m ohm ranges.

Test Current Accuracy

±0.

3%

Protection Fuses

Supply = 1.5A, HBC, 5 x 20mm, DIN
Current = 1A, HBC, 5 x 20mm, DIN
Voltage = 0.5A, HBC, 5 x 20mm, DIN

SAFETY

LVD BS EN 61010-1
EMC BS EN 50081-1

BS EN 50082-1

MECHANICAL
Case Height

110mm

Case Width

250mm

Case Depth

190mm

Bump Test

IEC68-2-29

Vibration Test

IEC1010, clause 8.3

Drop Test

IEC1010, clause 8.4

Impact Test

IEC1010, Clause 8.2

Weight

1.542kg

Rated environmental conditions
(1).Indoor Use
(2).Pollution Degree 2.
(3).Altitude up to 2000 meter.
(4).Relative humidity 80% max.
(5).Ambient temperature 0°C~40°C.